Influence of the ambient temperature to the sensor electronics

8)

A

verage temperature coefficient of the zero

signal (temperature error)

0.03 %/10 K

Influence of the superimposed gas and pressure to the accuracy

T

he spreading speed of the radar impulses in gas or vapour above the product is reduced by high

pressures. This effect depends on the superimposed gas or vapour and increases with low
temperatures. The following chart shows the deviation caused by some typical gases or vapours.
T

he stated values refer to the distance. Positive values mean that the measured distance is too

high, negative values that the measured distance is too small.

Gas phase

Temperature

1 bar/14.5 psig

10 bar/145 psig

50 bar/725 psig

A

ir/Nitrogen

20 °C/68 °F

0.00 %

0.22 %

1.2 %

A

ir/Nitrogen

200 °C/392 °F

0.00 %

0.13 %

0.74 %

H

ydrogen

20 °C/68 °F

-0.01 %

0.10 %

0.61 %

H

ydrogen

200 °C/392 °F

-0.02 %

0.05 %

0.37 %

W

ater (saturated

steam)

100 °C/212 °F

0.20 %

-

-

W

ater (saturated

steam)

180 °C/356 °F

-

2.1 %

-

Ambient conditions

A

mbient, storage and transport temperature

-40 … +70 °C (-40 … +158 °F)

Process conditions

F

or the process temperature and the vessel pressure, you also have to note the specifications on

the type label! Always the lowest value is applicable!

P

rocess temperature (measured on the process fitting depending on process seal

-

FKM (V

iton)

-40 … +80 °C (-40 … +176 °F)

-

FKM (V

iton) with temperature adapter

-40 … +150 °C (-40 … +302 °F)

-

FFKM (K

alrez 6375)

-20 … +150 °C (-4 … +302 °F)

-

graphite (coupling cone ceramic)

-60 … +250 °C (-76 … +482 °F)

-

graphite (coupling cone ceramic) with
temperature adapter

-60 … +400 °C (-76 … +752 °F)

V

essel pressure relating to the antenna cone

-

PTFE

-100 … 4000

kPa/-1 … 40 bar (-14.5 … 580 psi)

with PN 40

-

C

eramic

-100 … 16000

kPa/-1 … 160 bar (-14.5 … 2321 psi)

with PN 160

8)

R

elating to the nominal measuring range, in the temperature range

-40 … +80 °C .

VEGAPULS

66

standpipe version • 4 … 20 mA/HART four-wire

43