3 exchanging the electronics module, 4 software update – VEGA VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) Profibus PA User Manual

Page 45

Advertising
background image

?

E

041, E042, E043

l

Hardware error, electronics defective

à Exchange the instrument or send it in for repair

?

E

113

l

Communication conflict

à Exchange the instrument or send it in for repair

Depending on the reason for the fault and the measures taken, the
steps described in chapter "Set up" may have to be carried out again.

8.3 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.

In Ex applications only one instrument and one electronics module with
respective Ex approval may be used.

If there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.

The new electronics module must be loaded with the settings of the
sensor. These are the options:

l

At the factory by VEGA

l

Or on site by the user

In both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.

Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Electronics
module
").

The electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.

8.4 Software update

The software version of VEGAPULS 63 can be determined as follows:

l

on the type label of the electronics

l

via the indicating and adjustment module

l

via PACTware

You can view all software histories on our website www.vega.com.
Make use of this advantage and get registered for update information
via e-mail.

The following components are required to update the sensor software:

Reaction after fault rec-
tification

Sensor serial number

Assignment

VEGAPULS

63 • Profibus PA

45

8 Maintenance and fault rectification

28446

-EN

-120530

Advertising