VEGA VEGAPULS 62 (≥ 2.0.0 - ≤ 3.8) Profibus PA User Manual

Page 55

Advertising
background image

Input variable

Measured value

distance between process fitting and product sur-
face

Min. distance from antenna end

50 mm (1.969 in)

Recommended meas. range depending on the antenna diameter

-

ø

40 mm (1.575 in)

up to 10 m (32.81 ft)

-

ø

48 mm (1.89 in)

up to 15 m (49.21 ft)

-

ø

75 mm (2.953 in), ø 95 mm (3.74 in),

parabolic antenna

up to 30 m (98.43 ft)

Max. measuring range

up to 35 m (114.83 ft)

Reference conditions to measuring accuracy (similar to DIN EN 60770-1)

Reference conditions according to DIN EN 61298-1

-

Temperature

+

18 … +30 °C (+64 … +86 °F)

-

Relative humidity

45 … 75 %

-

Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Other reference conditions

-

Reflector

ideal reflector, e.g. metal plate 2 x 2 m

-

False reflections

Biggest false echo, 20 dB smaller than the useful
echo

Characteristics and performance data

Frequency

K-band (26 GHz technology)

Meas. interval approx.

1 s

Beam angle 3 dB

-

Encapculated antenna system

22°

-

Plastic horn antenna

10°

Step response or adjustment time

6)

<

1 s (dependent on the parameter settings)

Max. level change

Adjustable up to 1 m/min. (dependent on the
parameter adjustment)

Max. emitted HF power of the antenna system

-

Pulse peak power

<

2 mW

-

Pulse duration

<

2 ns

-

Average power

<

5 µW

-

Average power with 1 m distance

<

200 nW/cm

2

Max. emitted HF power of the antenna system - Version with increased sensitivity

-

Pulse peak power

<

10 mW

-

Pulse duration

<

2 ns

6)

Time to output the correct level (with max. 10 % deviation) after a sudden
level change.

VEGAPULS

62 • Profibus PA

55

10 Supplement

28445

-EN

-090305