4 mounting, 1 general instructions, 2 mounting preparations vegacal – VEGA VEGACAL series 60 User Manual

Page 6: 4mounting

Advertising
background image

4

Mounting

4

.1 General instructions

If the electronics module is defective, it can be replaced by the user.

In Ex applications only one instrument and one electronics module with
respective Ex approval may be used.

If there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.

The new electronics module must be loaded with the order and factory
data of the sensor. These are the options:

l

At the factory by VEGA

l

Or on site by the user

Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see "Setup").

In both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.

Caution:
The order data and company data contain important presettings for the
sensor. A reliable operation and a correction of the measurement is
not possible without these data.

Electronics modules are adapted to the respective sensor. They differ,
e.g. in the signal output, the power supply or approval.

First of all check by means of the overview in section Mounting
preparations, if you are using the suitable electronics module.
Compare the new electronics module with the existing one. The
specifications on the type label must correspond. This applies mainly
to instruments with approvals.

Warning:
Switch off voltage supply before starting the mounting procedure. The
replacement electronics must only be mounted in idle condition. Non-
observance will damage the electronics!

4

.2 Mounting preparations VEGACAL

Electronics module CL-E.60H. is suitable for VEGACAL - 4 … 20mA/
HART.

Electronics module CL-E.60P. is suitable for VEGACAL - Profibus PA.

Sensor serial number

Assignment

Electronics versions

6

Electronics module • VEGACAL series 60

4 Mounting

30531

-EN

-110411

Advertising