VEGA VEGAFLEX 65 4 … 20 mA_HART two-wire User Manual

Page 28

Advertising
background image

L

inearisation curve

Linear

E

nter the requested parameters via the appropriate keys, save your

settings and jump to the next menu item with the [->] key.

Caution:
N

ote the following if the VEGAFLEX 65 with corresponding approval is

used as part of an overfill protection system according to WHG (Water
R

esources Act):

I

f a linearisation curve is selected, the measuring signal is no longer

compulsorily linear proportional to the level. This must be taken into
consideration by the user, particularly when adjusting the switching
point on the level switch.

I

n this menu item you can enter an unambiguous designation for the

sensor, e.g. the measurement loop name or the tank or product
designation. In digital systems and in the documentation of larger
plants, a singular designation should be entered for exact identification
of individual measuring points.

S

ensor-TAG

Sensor

W

ith this menu item, the Basic adjustment is finished and you can now

jump to the main menu with the [ESC] key.

T

his function enables reading out parameter adjustment data as well

as writing parameter adjustment data into the sensor via the indicating
and adjustment module. A description of the function is available in the
operating instructions manual "Indicating and adjustment module".

T

he following data are read out or written with this function:

l

M

easured value presentation

l

A

djustment

l

M

edium

l

V

essel form

l

D

amping

l

L

inearisation curve

l

S

ensor-TAG

l

D

isplayed value

l

D

isplay unit

l

S

caling

l

C

urrent output

Sensor-TAG

Copy sensor data

28

VEGAFLEX

65 • 4 … 20

mA/HART - two-wire

6 S

et up with the indicating and adjustment module PLICSCOM

31845

-EN

-120329