VEGA VEGAFLEX 65 4 … 20 mA_HART two-wire User Manual

Page 44

Advertising
background image

-

2. HART

value (Secondary Value)

D

istance to the level - scaled (for example hl, %)

R

esolution, digital

>

1

mm (0.039 in)

Accuracy (according to DIN EN 60770-1)

P

rocess reference conditions according to DIN EN 61298-1

-

T

emperature

+

18 … +30 °C (+64 … +86 °F)

-

R

elative humidity

45 … 75 %

-

A

ir pressure

+

860 … +1060

mbar/+86 … +106 kPa

(+12.5 … +15.4

psig)

I

nstallation reference conditions

-

M

in. distance to installations

>

500

mm (19.69 in)

-

V

essel

metallic, ø 1 m (3.281 ft), centric installation,
process fitting flush with the vessel ceiling

-

M

edium

W

ater/Oil (dielectric figure ~2.0)

-

I

nstallation

P

robe end does not touch the vessel bottom

S

ensor parameter adjustment

G

ating out of false signals carried out

D

eviation

see diagrams
D

epending on the installation conditions, there can

be deviations which can be rectified with an
adaptation of the adjustment or a change of the
measured value offset in the DTM service mode.

L

0

-2 mm (-0.079

")

2 mm (0.079

")

-10 mm (0.394

")

5 mm (0.197

")

0,15 m (5.906

"

)

0,03 m (1.181

"

)

0,05 m

(1.969

")

1

Fig. 21: Deviation VEGAFLEX 65 in coax version in water

1

Dead zone - no measurement possible in this area

L

Probe length

44

VEGAFLEX

65 • 4 … 20

mA/HART - two-wire

10 S

upplement

31845

-EN

-120329

Advertising