4 exchanging the electronics module – VEGA VEGAFLEX 61 4 … 20 mA_HART two-wire User Manual

Page 44

Advertising
background image

3

C

able: Pull the cable out of the gravity weight

4

S

horten the cable/rod with a cut-off wheel or metal saw at the

lower end. Make sure the length is correct.

5

C

able: cable protrudes approx. 40 mm (1.575 in) into the gravity

weight.

6

C

able: Fasten the cable with three pins, torque 7 Nm (5.16 lbf ft)

7

E

nter new probe length and then carry out a fresh adjustment (see

"S

etup procedure, Carrying out min. adjustment - Carrying out

max. adjustment").

Fig. 26: Shortening the cable probe

8

.4 Exchanging the electronics module

I

f the electronics module is defective, it can be replaced by the user.

I

n Ex applications only one instrument and one electronics module with

respective Ex approval may be used.

I

f there is no electronics module available on site, one can be ordered

from the VEGA agency serving you.

T

he new electronics module must be loaded with the settings of the

sensor. These are the options:

l

A

t the factory by VEGA

l

O

r on site by the user

I

n both cases, the sensor serial number is necessary. The serial

numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.

Information:
W

hen loading on site, first of all the order data must be downloaded

from the Internet (see operating instructions manual "Oscillator").

Sensor serial number

44

VEGAFLEX

61 • 4 … 20

mA/HART two-wire

8 M

aintenance and fault rectification

31833

-EN