VEGA VEGAFLEX 61 4 … 20 mA_HART two-wire User Manual

Page 54

Advertising
background image

L

3 mm

(0.118

")

0

-3 mm

(-0.118

")

15 mm

(0.591

")

20 mm

(0.787”)

0,08 m

(3.15

")

0,3 m

(11.811

")

20 m

(787.4

")

1

Fig. 32: Deviation VEGAFLEX 61 in cable version, probe length L > 20 m in water
1

Dead zone - no measurement possible in this area

L

Probe length

L

0

20 mm

(0.787

")

0,25 m

(9.843

")

5 mm

(0.197

")

-5 mm

(-0.197

")

15 mm

(0.591

"

)

0,3 m

(11.811

")

20 m

(787.4

")

0,15 m

(5.906

"

)

1

1

Fig. 33: Deviation VEGAFLEX 61 in cable version, probe length L > 20 m in oil
1

Dead zone - no measurement possible in this area

L

Probe length

Influence of the ambient temperature to the sensor electronics

T

emperature drift

0.03 %/10 K

relating to the max. measuring range

or max. 0.3 %

T

emperature drift - Digital output

3

mm/10 K relating to the max. measuring range or

max. 10 mm

54

VEGAFLEX

61 • 4 … 20

mA/HART two-wire

10 S

upplement

31833

-EN

-100426