VEGA VEGABAR 63 Foundation Fieldbus User Manual

Page 59

Advertising
background image

Adjustment elements

4 keys

Protection class

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

Materials

-

Housing

ABS

-

Inspection window

Polyester foil

Power supply

Operating voltage

-

Non-Ex instrument

9 … 32 V DC

-

EEx-ia instrument

9 … 24 V DC

-

EEx-d instrument

9 … 32 V DC

Operating voltage with lighted indicating and adjustment module

23)

-

Non-Ex instrument

13 … 32 V DC

-

EEx-ia instrument

13 … 24 V DC

-

EEx-d instrument

13 … 32 V DC

Power supply by/max. number of sensors

-

H

1 power supply

max. 32 (max. 10 with Ex)

Electrical protective measures

Protection class

-

Housing, standard

IP

66/IP 67

24)

-

Aluminium and stainless housing (op-
tionally available)

IP

68 (1 bar)

25)

-

Transmitter in IP 68 version

IP

68

-

External housing

IP

65

Overvoltage category

III

Protection class

II

Approvals

Depending on the version, instruments with approvals can have different technical data.

23)

Is available at a later date for instruments with StEx, WHG or ship approval
as well as country-specific approvals such as those according to FM or
CSA.

24)

Instruments with gauge pressure measuring ranges cannot detect the am-
bient pressure when submerged, e.g. in water. This can lead to falsification
of the measured value.

25)

Only with instruments with absolute pressure ranges.

VEGABAR

63 • Foundation Fieldbus

59

10 Supplement

32464

-

EN

-081209

Advertising