VEGA VEGAFLEX 67 (-200…+400°C) Foundation Fieldbus User Manual
Page 48
2
1
-100°C
(-148°F)
-200°C
(-328°F)
100 bar
(1450 psig)
200 bar
(2900 psig)
300 bar
(4350 psig)
400 bar
(5800 psig)
380 bar
(5510 psig)
350 bar
(5070 psig)
0°C
(32°F)
200°C
(392°F)
100°C
(212°F)
400°C
(752°F)
300°C
(572°F)
Fig. 25: High temperature version: Process pressure - process temperature
1
Process temperature
2
Process pressure
Process temperature (thread or flange tem-
perature)
-
200 … +400 °C (-328 … +752 °F)
The measurement error from the process condi-
tions is in the specified pressure and temperature
range of below 1 %.
1
2
-110°C
(-166°F)
-200°C
(-328°F)
-20°C
(-4°F)
0°C
(32°F)
-40°C
(-40°F)
60°C
(140°F)
80°C
(176°F)
100°C
(212°F)
200°C
(392°F)
100°C
(212°F)
400°C
(752°F)
300°C
(572°F)
Fig. 26: Ambient temperature - process temperature (version -200 … +400 °C/-328 … +752 °F)
1
Ambient temperature
2
Process temperature (graphite seal)
48
VEGAFLEX
67 • Foundation Fieldbus
10 Supplement
32314
-EN
-120329