4 mounting – sensor housing, 4 mounting - sensor housing – VEGA VEGAFLEX series 60 - External housing User Manual
Page 8
90mm (3
35
/
64
")
R3,5mm
(
9
/
64
")
3mm
(
1
/
8
")
70mm (2
3
/
4
")
8mm
(
5
/
16
")
93mm (3
21
/
32
")
110mm (4
21
/
64
")
Fig. 2: Drilling template - wall mounting plate
Tip:
Mount the wall mounting plate so that the cable entry of the
socket housing points downward. Rain and condensation
water can thus drain off. The socket housing can be displaced
by 180° to the wall mounting plate.
Turn the cable gland of the instrument housing downward. The
basic body of the instrument housing can be turned by 330°
without any tools.
Warning:
The four screws of the socket housing must only be hand-
screwed. A torque > 5 Nm (3.688 lbf ft) can damage the wall
mounting plate.
4
.4 Mounting - sensor housing
à
Mount the sensor to the bottom, depending on the process
fitting
Note:
When mounting the sensor, note the instructions in the
operating instructions manual of the sensor.
8
VEGAFLEX - External housing
Mounting
34106
-EN
-080227