VEGA VEGAFLEX 65 Profibus PA User Manual

Page 43

Advertising
background image

L

0

-2 mm (-0.079

")

2 mm (0.079

")

10 mm (0.394

")

5 mm (0.197

")

-5 mm (-0.197

")

0,15 m (5.906

"

)

0,1 m (3.937

"

)

0,15 m

(5.906

")

0,1 m

(3.937

")

1

1

Fig. 18: Deviation VEGAFLEX 65 in coax version in oil

1

Dead zone - no measurement possible in this area

L

Probe length

Influence of the ambient temperature to the sensor electronics

Temperature drift

0.03 %/10 K relating to the max. measuring range
or max. 0.3 %

Temperature drift - Digital output

3 mm/10 K relating to the max. measuring range or
max. 10 mm

Ambient conditions

Ambient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

Process pressure

-1 … +40 bar/-100 … +4000 kPa (-
-14.5 … +580 psig), depending on the process
fitting

Process temperature (thread or flange tem-
perature)

-

FKM (Viton)

-40 … +150 °C (-40 … +302 °F)

-

EPDM

-40 … +150 °C (-40 … +302 °F)

-

FFKM (Kalrez 6375)

-20 … +150 °C (-4 … +302 °F)

The measurement error from the process condi-
tions is in the specified pressure and temperature
range of below 1 %.

VEGAFLEX

65 • Profibus PA

43

10 Supplement

31847

-

EN

-120329

Advertising