8 Maintenance and fault rectification

VEGABAR 67 • Profibus PA - climate compensated

39445-EN-130321

Error code Cause

Rectification

E113

Communication con-

flict

– Exchange the instrument or send it

in for repair

8.3 Calculation of total deviation (according to

DIN 16086)

The total deviation F

total

according to DIN 16086 is the sum of basic

accuracy F

perf

and longterm stability F

stab

. F

total

is also called maximum

practical deviation or utility error.
F

total

= F

perf

+ F

stab

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

)

With:

F

total

: Total deviation

F

perf

: Basic accuracy

F

stab

: Long-term drift

F

T

: Temperature coefficient (influence of medium or ambient

temperature)

F

Kl

: Deviation

Level measurement 1.5 mWs
Ambient temperature 40 °C, product temperature 10 °C
VEGABAR 67 with measuring range 0.2 bar
Deviation < 0.1 %
Calculation of the set Turn Down:
TD = 200 mbar/147 mbar, TD = 1.4
Basic accuracy digital output signal in percent:

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

)

F

T

= (0.15 + 0.4 x TD) %

F

Kl

= 0.1 %

F

perf

= √((0.15 % + 0.4 % x 1.4)

2

+ (0.1 %)

2

)

F

perf

= 0.72 %

Total deviation digital output signal in percent:

F

total

= F

perf

+ F

stab

F

stab

= (1 % x 1.4)/year

F

stab

= 1.4 %

F

total

= 0.72 % + 1.4 % = 2.12 %

Total deviation digital output signal absolute:

F

total

= 2.12 % x 147 mbar/100 % = 3.11 mbar = 31.77 mm

8.4 Exchanging the electronics module

In case of a defect, the electronics module can be exchanged by the

user against an identical type. If no electronics module is available on

side, the module can be ordered from the agency serving you.

Total deviation

Example