VEGA VEGAPULS SR 68 4 … 20 mA_HART - four-wire User Manual

Page 33

Advertising
background image

33

5 Connecting to power supply

VEGAPULS SR 68 • 4 … 20 mA/HART - four-wire

38295-EN-130624

5 6 7 8

4...20mA

2

1

2

+

( )

(-)

1

Fig. 28: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment

2 For display and adjustment module or interface adapter

Information:

The connection of an external display and adjustment unit is not pos-

sible with this double chamber housing.

power supply

4...20mA

active

passive

common

IS GND

5

1

/L

/N

PE

2

+

( )

(-)

6 7

8

Fig. 29: Connection compartment with double chamber housing with mains

voltage

Terminal

Function

Polarity

1

Voltage supply

+/L

2

Voltage supply

-/N

5

4 … 20 mA output (active)

+

6

4 … 20 mA output (passive)

+

7

Mass - output

-

8

Function ground with installa-

tion according to CSA

Electronics compartment

Connection compartment

with mains voltage

Advertising