VEGA VEGABAR 55 4 … 20 mA_HART - climate compensated User Manual

Page 30

Advertising
background image

30

6 Set up with the display and adjustment module PLICSCOM

VEGABAR 55 • 4 … 20 mA/HART - climate compensated

39441-EN-130321

Information:

The steps 1, 3 and 4 are not necessary for instruments which are

already preset according to customer specifications!
You can find the data on the type label on the instrument or in the

menu items of the min./max. adjustment.
The display and adjustment module enables the adjustment without

filling or pressure. Thanks to this, you can carry out your settings

already in the workshop without the instrument having to be installed.
The actual measured value is also displayed in the menu items for

min./max. adjustment.

In this menu item you select the adjustment unit as well as the unit for

the temperature indication in the display.
To select the adjustment unit (in the example switching over from bar

to mbar), proceed as follows:

2)

1. Push the [OK] button in the measured value display, the menu

overview is displayed.

Basic adjustment
Display
Diagnostics
Service
Info

2. Confirm the menu "Basic adjustment" with [OK], the menu item

"Unit" will be displayed.

Unit
Unit of measurement
bar

Temperature unit
°C

3. Activate the selection with [OK] and select "Units of measure-

ment with [->].

4. Activate the selection with [OK] and select the requested unit with

[->] (in the example mbar).

5. Confirm with [OK] and move to position correction with [->].
The adjustment unit is thus switched over from bar to mbar.
Information:

When switching over to adjustment in a height unit (in the example

from bar to m), the density also has to be entered.
Proceed as follows:
1. Push the [OK] button in the measured value display, the menu

overview is displayed.

2. Confirm the menu "Basic adjustment" with [OK], the menu item

"Units of measurement" will be displayed.

3. Activate the selection with [OK] and select the requested unit with

[->] (in the example m).

Select unit

2)

Selection options: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, inH

2

O,

mmH

2

O.

Advertising