Page 58

Advertising
background image

58

10 Supplement

VEGABAR 55 • 4 … 20 mA/HART - climate compensated

39441-EN-130321

Display and adjustment module
Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the sensor without cover IP 40

Material

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Voltage supply
Operating voltage

Ʋ Non-Ex instrument

12 … 36 V DC

Ʋ Ex-ia instrument

12 … 30 V DC

Ʋ Exd instrument

18 … 36 V DC

Operating voltage with illuminated display and adjustment module

Ʋ Non-Ex instrument

20 … 36 V DC

Ʋ Ex-ia instrument

20 … 30 V DC

Ʋ Ex d instrument

20 … 36 V DC

Permissible residual ripple

Ʋ < 100 Hz

U

ss

< 1 V

Ʋ 100 Hz … 10 kHz

U

ss

< 10 mV

Load

see diagram

1000

750

500

250

12

18

16

14

20

22

24

26

28

30

32

34

36

V

4

1

2

3

Fig. 29: Voltage diagram
1 HART load

2 Voltage limit Ex-ia instrument

3 Voltage limit non-Ex/Ex-d instrument

4 Operating voltage