VEGA VEGABAR 55 4 … 20 mA_HART - climate compensated User Manual

Page 46

Advertising
background image

46

8 Maintenance and fault rectification

VEGABAR 55 • 4 … 20 mA/HART - climate compensated

39441-EN-130321

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

+ (F

a

)

2

)

With:

F

total

: Total deviation

F

perf

: Basic accuracy

F

stab

: Long-term drift

F

T

: Temperature coefficient (influence of medium or ambient

temperature)

F

Kl

: Deviation

F

a

: Error current output

Example
Pressure measurement in the pipeline 8 bar (800 KPa)
Ambient temperature 40 °C, product temperature 10 °C, hence within

the compensated range
VEGABAR 55 climate compensated, with measuring range 10 bar
Calculation of the set Turn Down: TD = 25 bar/8 bar, TD = 1.25
Basic accuracy digital output signal in percent:

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

)

F

T

= (0.05 + 0.1 x TD)%

F

Kl

= 0.25 %

F

perf

= √((0.05 + 0.1 x 1.25)%)

2

+ (0.25 %)

2

)%

F

perf

= 0.31 %

Total deviation digital output signal in percent:

F

total

= F

perf

+ F

stab

F

stab

= (0.1 % x TD)/year

F

stab

= (0.1 % x 1.25)/year

F

stab

= 0.125 %

F

total

= 0.31 % + 0.125 % = 0.43 %

Total deviation digital output signal absolute:

F

total

= 0.43 % x 8 bar/100 % = 34.4 mbar

Basic accuracy analogue output signal in percent:

F

perf

= √((F

T

)

2

+ (F

Kl

)

2

+ (F

a

)

2

)

F

T

= (0.05 + 0.1 x TD)%

F

Kl

= 0.25 %

F

a

= 0.15 %

F

perf

= √((0.05 + 0.1 x 1.25)%)

2

+ (0.25 %)

2

+ (0.15%)

2

)

F

perf

= 0.34 %

Total deviation analogue output signal in percent:

F

total

= F

perf

+ F

stab

F

stab

= (0.1 % x TD)/year

F

stab

= (0.1 % x 1.25)/year

F

stab

= 0.125 %

F

total

= 0.34 % + 0.125 % = 0.47 %

Advertising