2 principle of operation – VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) HART and accumulator pack User Manual

Page 11

Advertising
background image

11

3 Product description

VEGAPULS 62 • HART and accumulator pack

40793-EN-121011

The instrument is available in two different electronics versions. Each

version can be identified via the product code on the type label as

well as on the electronics.

Standard electronics type PS60HK.-

Electronics with increased sensitivity type PS60HS.-

The scope of delivery encompasses:

Radar sensor with integrated accumulator

Battery charger

Documentation

– this operating instructions manual

– Test certificate measuring accuracy (optional)

– Operating instructions manual "Indicating and adjustment

module" (optional)

– Ex-specific "Safety instructions" (with Ex versions)

– if necessary, further certificates

3.2 Principle of operation

The VEGAPULS 62 radar sensor can be used in a wide variety of ap-

plications for continuous level measurement of liquids. It is suitable for

applications in storage vessels, reactors and process vessels, even

under extremely difficult process conditions.
Dependent on the application range, different versions are used:

Antenna

– Small tanks and process vessels, measurement of virtually all

products: Horn antenna ø 40 mm

– Storage tanks and process vessels, measurement of products

such as solvents, hydrocarbons and fuels under extremely dif-

ficult process conditions: Horn antenna ø 48 … 95 mm

– Products with low ε

r

value with large measuring distances:

Parabolic antenna

Electronics

– Products with an ε

r

value ≥1.8: Standard electronics

– Products with an ε

r

value < 1.8, ≥1.5; applications with very bad

reflective properties: Electronics with increased sensitivity

The actual values that can be reached depend on the measurement

conditions, the antenna system or the standpipe or bypass.

The antenna of the radar sensor emits short radar pulses with a

duration of approx. 1 ns. These pulses are reflected by the product

and received by the antenna as echoes. The transit time of the radar

pulses from emission to reception is proportional to the distance and

hence to the level. The determined level is converted into an appropri-

ate output signal and outputted as measured value.

The instrument is powered via an integrated accumulator. This allows

operation of the instrument independently of the mains or a control

system. To achieve a longer battery lifetime, the instrument is preset

to HART multidrop. This puts a load of only 4 mA current on the ac-

cumulator.