VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) HART and accumulator pack User Manual

Page 38

Advertising
background image

38

6 Set up with the indicating and adjustment module

VEGAPULS 62 • HART and accumulator pack

40793-EN-121011

Since a radar sensor is a distance measuring instrument, the distance

from the sensor to the product surface is measured. For indication of

the real level, an allocation of the measured distance to the percent-

age height must be carried out.
To perform the adjustment, enter the distance with full and empty ves-

sel, see the following example:

100%

0%

0,5 m (19.68

")

5 m

(196.9

")

2

1

Fig. 30: Parameter adjustment example min./max. adjustment
1 Min. level = max. meas. distance

2 Max. level = min. meas. distance

If these values are not known, an adjustment with the distances of for

example 10 % and 90 % is possible. Starting point for these distance

specifications is always the seal surface of the thread or flange. By

means of these settings, the real level will be calculated.
The real product level during this adjustment is not important, be-

cause the min./max. adjustment is always carried out without chang-

ing the product level. These settings can be made ahead of time

without the instrument having to be installed.

Proceed as follows:
1. Select the menu item "Setup" with [->] and confirm with [OK].

Now select with [->] the menu item "Min. adjustment" and confirm

with [OK].

2. Edit the percentage value with [OK] and set the cursor to the

requested position with [->].

Setup/Adjustment

Setup/Min. adjustment

Advertising