VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) HART and accumulator pack User Manual

Page 64

Advertising
background image

64

11 Supplement

VEGAPULS 62 • HART and accumulator pack

40793-EN-121011

Process conditions
For the process conditions, please also note the specifications on the type label. The lowest value

always applies.

Seal

Antenna impedance

cone

Process temperature (measured on the process

fitting)

FKM (SHS FPM 70C3

GLT)

PTFE

-40 … +130 °C (-40 … +266 °F)

PTFE

4)

-40 … +200 °C (-40 … +392 °F)

PEEK

5)

-40 … +200 °C (-40 … +392 °F)

FFKM (Kalrez 6375)

PTFE

-20 … +130 °C (-4 … +266 °F)

PEEK

-20 … +250 °C (-4 … +482 °F)

FFKM (Kalrez 2035)

PTFE

-15 … +130 °C (5 … +266 °F)

PEEK

-15 … +210 °C (5 … +410 °F)

FFKM (Kalrez 6230)

PTFE

-15 … +130 °C (5 … +266 °F)

PEEK

-15 … +250 °C (5 … +482 °F)

Graphite

Ceramic

-200 … +450 °C (-328 … +842 °F)

Graphite (process fitting

Hastelloy C 22)

Ceramic

-196 … +400 °C (-321 … +752 °F)

Vessel pressure - horn antenna

Ʋ Antenna impedance cone PTFE

-1 … 40 bar (-100 … 4000 kPa/-14.5 … 580 psig)

Ʋ Antenna impedance cone PP

-1 … 3 bar (-100 … 300 kPa/-14.5 … 43.5 psig)

Ʋ Antenna impedance cone PEEK

-1 … 100 bar (-100 … 10000 kPa/-14.5 … 1450 psig)

Ʋ Antenna impedance cone ceramic

-1 … 160 bar (-100 … 16000 kPa/-14.5 … 2320 psig)

Vessel pressure - parabolic antenna

-1 … 6 bar (-100 … 6000 kPa/-14.5 … 870 psig)

Vessel pressure with swivelling holder

-1 … 1 bar (-100 … 100 kPa/-14.5 … 14.5 psig)

Vessel pressure relating to the flange

nominal stage

see supplementary instructions manual "Flanges ac-

cording to DIN-EN-ASME-JIS"

Vibration resistance

Ʋ Horn antenna

4 g at 5 … 200 Hz according to EN60068-2-6 (vibration

with resonance)

Ʋ Parabolic antenna

1 g at 5 … 200 Hz according to EN60068-2-6 (vibration

with resonance)

Shock resistance

Ʋ Horn antenna

100 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

Ʋ Parabolic antenna

25 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

Data on rinsing air connection
Pressure max.

6 bar (87.02 psig)

Air quantity without reflux valve, with

Ʋ 0.2 bar (2.9 psig)

2.0 m

3

/h

4)

Not with steam

5)

Not with steam

Advertising