3 Planning instructions for isolating systems

Chemical seal CSB • for VEGADIF 65

36133-EN-130417

this, an additional vacuum load from the filling oil in the capillaries is

avoided.

+ –

+

Fig. 11: Preferred mounting below the lower chemical seal

= ∆p

h g

+

h

1

Fig. 12: Density measurement with VEGADIF 65, h = defined mounting dis-

tance, Δp = differential pressure, ρ = density of the medium, g = acceleration of

gravity
1 VEGADIF 65

1

+

h

2

3

Fig. 13: Interface measurement with VEGADIF 65
1 VEGADIF 65

2 Liquid with higher density

3 Liquid with lower density

Density measurement

Interface measurement