7 Set up with the display and adjustment module

MINITRAC 31 • 4 … 20 mA/HART - four-wire

43389-EN-131119

You can enter the adjustment point (ct/s) manually.

You can let the adjustment point be determined by MINITRAC 31.

This menu item appears only if you have selected under adjustment

mode (setup/adjustment mode) the "Two-point adjustment".
In this menu item you can set the max. pulse rate (ct/s) at which the

sensor should switch over.
Empty the vessel until the MINITRAC 31 is uncovered.
You thus get the max. pulse rate (ct/s) for the "uncovered" adjustment.
Enter the requested pulse rate manually or let the rate be determined

by MINITRAC 31. Automatic determination of the pulse rate should be

given preference.

You can enter the adjustment point (ct/s) manually.

You can let the adjustment point be determined by MINITRAC 31.

In this menu item you can activate or deactivate the current output.
The MINITRAC 31 checks if with activated current output there is re-

ally an instrument connected.
If no instrument is connected to the current output, you have to deacti-

vate the current output.

Adjustment "uncovered"

(two-point adjustment)

Setup/Current output