7 Set up with the display and adjustment module

SOLITRAC 31 • 4 … 20 mA/HART - four-wire

43387-EN-131119

7 Set up with the display and adjustment

module

7.1 Insert display and adjustment module

The display and adjustment module can be inserted into the sensor

and removed again at any time. It is not necessary to interrupt the

power supply.
Proceed as follows:
1. Unscrew the small housing cover
2. Place the display and adjustment module in the desired position

on the electronics (you can choose any one of four different posi-

tions - each displaced by 90°)

3. Press the display and adjustment module onto the electronics

and turn it to the right until it snaps in.

4. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The display and adjustment module is powered by the sensor, an ad-

ditional connection is not necessary.

1

2

Fig. 16: Insert display and adjustment module

Note:

If you intend to retrofit the instrument with a display and adjustment

module for continuous measured value indication, a higher cover with

an inspection glass is required.