11 supplement, 1 technical data – VEGA SOLITRAC 31 4 … 20 mA_HART - four-wire With SIL qualification User Manual

Page 72

Advertising
background image

72

11 Supplement

SOLITRAC 31 • 4 … 20 mA/HART - four-wire

43387-EN-131119

11 Supplement

11.1 Technical data

General data
316L corresponds to 1.4404 or 1.4435
Materials, non-wetted parts

Ʋ Detector tube

316L

Ʋ Scintillation material

PVT (Polyvinyltoluene)

Ʋ Aluminium die-casting housing

Aluminium die-casting AlSi10Mg, powder-coated - basis:

Polyester

Ʋ Stainless steel housing

316L

Ʋ Seal between housing and housing

cover

NBR (stainless steel housing, investment casting),

silicone (Aluminium housing)

Ʋ Inspection window in housing cover

(optional)

Polycarbonate

Ʋ Ground terminal

316L

Ʋ Mounting accessory

316L

Process fittings

Ʋ Fastening lugs

ø 9 mm (0.35 in), hole centre distance 119 mm (4.69 in)

Weight

Ʋ Aluminium housing, with electronics

3.4 kg (7.5 lbs) + measurement tube

Ʋ Stainless steel housing, with electron-

ics

8.36 kg (18.43 lbs) + measurement tube

Ʋ Measurement tube

7.1 kg/m (4.77 lbs/ft)

Max. torque, mounting screws - fastening

lugs on the sensor housing

50 Nm (36.88 lbf ft)

Max. torque for NPT cable glands and Conduit tubes

Ʋ Aluminium/Stainless steel housing

50 Nm (36.88 lbf ft)

Input variable
Measured variable

The measured variable is the intensity of the gamma

radiation of an isotope. When the radiation intensity

decreases, for example due to rising medium, the

measured value of SOLITRAC 31 changes in proportion

to the level.

Advertising