VEGA VEGAPULS 66 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol User Manual

Page 61

Advertising
background image

61

11 Supplement

VEGAPULS 66 • Modbus and Levelmaster protocol

41366-EN-121011

Ʋ Closing cap

M20 x 1.5; ½ NPT

Plug options

Ʋ Supply and signal circuit (only with low

voltage)

M12 x 1

Ʋ Indication circuit

M12 x 1

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

Indicating and adjustment module
Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Measured value indication

Ʋ Number of digits

5

Ʋ Size of digits

W x H = 7 x 13 mm

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the sensor without cover IP 40

Materials

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Integrated clock
Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Measurement electronics temerature
Resolution

1 °C (1.8 °F)

Accuracy

±1 °C (1.8 °F)

Voltage supply
Operating voltage

8 … 30 V DC

Interpolation protection

Available

Electrical protective measures
Protection, depending on housing version

Ʋ Plastic housing

IP 66/IP 67

Ʋ Aluminium housing, stainless steel

housing - investment casting, stain-

less steel housing - electro-polished

IP 66/IP 68 (0.2 bar)

4)

4)

The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.