9 Diagnostics and service

VEGAFLEX 86 • Modbus and Levelmaster protocol

41848-EN-130314

Information:

Please keep the stated torque so that the max. tensile strength

remains.

The rod or cable of the probe can be shortened individually.

1. Mark the requested length with mounted measuring rod.
2. Cable: Loosen the pins on the gravity weight (hexagon 3)
3. Cable: remove the pins
4. Cable: Pull the cable out of the gravity weight
5. Shorten the cable/rod with a cut-off wheel or metal saw at the

marking. Take note of the specifications in the following illustration

when shortening the cable.

6. Cable with gravity weight: Shift the cable according to the drawing

into the gravity weight

7. Cable with gravity weight: Fasten cable with the pins, torque 7 Nm

(5.16 lbf ft)
Cable with centering weight: Fasten cable with the pins, torque

7 Nm (5.16 lbf ft) and fix the clamping part on the centering

weight.

8. Enter new probe length and then carry out a fresh adjustment

(see "Setup procedure, Carrying out min. adjustment - Carrying

out max. adjustment").

4

1

1

2

2

3

A

B

C

45 mm (1.77")

100 mm

(3.94")

25 mm (0.98")

50 mm

(1.97")

Fig. 38: Shortening the cable probe
A Gravity weight - cable ø 4 mm

B Gravity weight - cable ø 2 mm

C Centering weight - cable ø 2 mm

1 Pins

2 Thread M8 for eye-bolt

3 Fixing screw - centering weight

9.7 Software update

The following components are required to update the sensor soft-

ware:

Sensor

Voltage supply

Shorten cable/rod