3 spacers - plastic (open on the side) (k4, k5) – VEGA VEGAFLEX series 80 Centering User Manual

Page 12

Advertising
background image

12

2 Mounting

Centering • for VEGAFLEX series 80

44967-EN-131119

The retaining rings fix the spacer in the desired position. A certain

clearance is permitted.

Note:

The retaining rings can be moved only in one direction. If you have

already passed the desired position, cut the retaining ring with a side

cutter. Use a new retaining ring.

2.3 Spacers - Plastic (open on the side) (K4, K5)

1

Fig. 9: Mounting the spacer (open on the side)
1 Spacer (K4, K5) for tube inside diameter ≤ 100 mm (4")

1. Measure the diameter of the standpipe or bypass tube and adapt

the spacers accordingly. The notches on the 4-arm spacer allow

simple trimming down to tube diameters DN 50 (2") and DN 80

(3").
For cutting, use a metal saw or a pair of strong pincers or side

cutters.

2. With rod versions of the probe, the spacer should be mounted at

the lower end of the probe rod. Use as few spacers as possible.

With rod probes, one spacer at the probe end is usually sufficient.

If several spacers are required, they should be mounted large

distances apart. Determine where a spacer should be placed and

mark the position.
With cable version, the spacers should be mounted at a min.

distance of approx. 1.50 m (5 ft).

3. Plug the spacer (1) from the side onto the rod or cable of the

probe.