VEGA VEGAFLEX series 80 Centering User Manual

Page 16

Advertising
background image

16

2 Mounting

Centering • for VEGAFLEX series 80

44967-EN-131119

1

4

2

1

4

2

Fig. 13: Mounting the spacer (K3 or M1, M2) on the gravity weight (S1, S2)
1 Gravity weight (S1, S2)

2 Spacer - Plastic (K3) for tube inside diameter ≤ 100 mm (4")

3 Spacers - Metal (M1, M2)

4 Retaining ring

1. Measure the diameter of the standpipe or bypass tube and adapt

the spacer (2) accordingly. The notches on the 4-arm spacer (2)

allow simple trimming down to tube diameters DN 50 (2") and

DN 80 (3").
For cutting, use a metal saw or a pair of strong pincers or side

cutters.
With spacers of metal (3), you have to use a spacer that fits.

2. Plug the spacer (2 or 3) onto the short neck of the gravity weight

(1)

3. Slide a retaining ring (4) onto the short neck of the gravity weight

(1)

Advertising