54

6 Set up with the display and adjustment module

MINITRAC 31 • Foundation Fieldbus

41782-EN-131119

Menu

Menu item

Default value

Display

Language

Selected language

Displayed value

Pulse rate

Display unit

ct/s

With this function

Load parameter adjustment data from the sensor into the display

and adjustment module

Write parameter adjustment data from the display and adjustment

module into the sensor

This parameter is described in the operating instructions manual

"Display and adjustment module".
Info
In this menu you will find the following menu items:

Instrument name - shows instrument name and serial number

Instrument version - shows hardware and software version of the

instrument

Date of manufacture - shows calibration date and the date of the

last change

Device ID - shows the device ID and the sensor TAG (PD_TAG)

Instrument features - shows further instrument features

These parameters are described in the operating instructions manual

"Display and adjustment module".

6.6 Parameter adjustment - X-ray alarm

X-ray radiation, for example, cause with weld joint tests can influence

your measured values unnoticeably. Hence it is importnat to recog-

nized interfering X-ray radiation.
The MINITRAC 31 recognizes the X-ray raditation and reacts from an

individually definable raditiation volume.
The level or density measuring instrument acts as Master and the

MINITRAC 31 operates as Slave.
First of all, define the function of the Slave instrument before you

define the Master instrument. The Master instrument can thus imme-

diately recognize the connected Slave.
For this purpose, the Slave instrument must be defined as "X-ray

alarm". Select under the menu item "Setup/Application" the function

"X-ray alarm".
You can set the address (MGC) on the Slave instrument at will. Only

the address "0 - 0" is reserved for the Master instrument.
Set the address setting (MGC) on the Master instrument to "0 - 0".

Additional adjustments -

Copy instrument settings

Info

X-ray alarm