VEGA VEGAFLEX 83 4 … 20 mA_HART two-wire Rod probe - Polished version With SIL qualification User Manual

Page 39

Advertising
background image

39

7 Set up with the display and adjustment module

VEGAFLEX 83 • 4 … 20 mA/HART two-wire

44225-EN-130823

Note:

The selection of the application has a considerable influence on all

other menu items. Keep in mind that as you continue with the param-

eter adjustment, individual menu items are only optionally available.
You have the option of choosing the demonstration mode. In this

mode, the sensor ignores the parameters of the application and

reacts immediately to any change.

This mode is only suitable for test and demonstration purposes and

must not be used in a safety-instrumented application (SIL).

In this menu item, you can define the medium type (product).
This menu item is only available if you have selected level measure-

ment under the menu item "Application".

You can choose between the following medium types:

Dielectric figure Medium type

Examples

> 10

Water-based liq-

uids

Acids, alcalis, water

3 … 10

Chemical mix-

tures

Chlorobenzene, nitro lacquer, aniline,

isocyanate, chloroform

< 3

Hydrocarbons

Solvents, oils, liquid gas

This menu item is only available, if you have chosen interface meas-

urement under the menu item "Application". In this menu item you can

enter if there is a superimposed gas phase in your application.
Only set the function to "Yes", if the gas phase is permanently pre-

sent.

This menu item is only available if you have selected interface meas-

urement under the menu item "Application". In this menu item you can

choose the medium type of the upper medium.

Setup - Application - Me-

dium, dielectric constant

Setup - Application - Gas

phase

Setup - Application - Di-

electric constant

Advertising