VEGA VEGASWING 66 - Relay User Manual

Page 9

Advertising
background image

9

3 Product description

VEGASWING 66 • - Relay

43756-EN-130322

3.2 Principle of operation

VEGASWING 66 is a point level sensor with tuning fork for level

detection.
It is designed for industrial use in all areas of process technology and

can be used in liquids. It is particularly suitable for applications with

high temperatures up to 450 °C (842 °F) and high process pressure

up to 160 bar (2320 psig).
Typical applications are overfill and dry run protection. The small

tuning fork allows use in pipelines. Thanks to its simple and rugged

measuring system, VEGASWING 66 is virtually unaffected by the

chemical and physical properties of the liquid.
It functions even under difficult conditions such as turbulence, foam

generation, buildup, strong external vibration or changing products.
Function monitoring

The electronics module of VEGASWING 66 continuously monitors via

frequency evaluation the following criteria:

Strong corrosion or damage on the tuning fork

Loss of vibration

Line break to the vibrating drive

If a malfunction is detected or in case of power failure, the electron-

ics takes on a defined switching condition, i.e. the relay deenergises

(safe condition).

The tuning fork vibrates at its mechanical resonance frequency of

approx. 1400 Hz. When the tuning fork is submerged in the product,

the frequency changes. This change is detected by the integrated

electronics module and converted into a switching command.

VEGASWING 66 can be operated without external evaluation system.

The integrated electronics evaluates the level signal and outputs a

switching signal. With this switching signal, a connected device can

be operated directly (e.g. a warning system, a pump etc.).
The data for power supply are specified in chapter "Technical data".

3.3 Operation

With the factory setting, products with a density > 0.07 g/cm³

(0.025 lbs/in³) can be detected. The instrument can be adapted to

products with lower density.
On the electronics module you will find the following display and

adjustment elements:

Signal lamp for indication of the operating condition (green)

Signal lamp for indication of the switching condition (yellow)

Control lamp for fault indication (red)

DIL switch for sensitivity adjustment

Mode switch for selecting the switching behaviour (min./max.)

Application area

Functional principle

Voltage supply

Advertising