VEGA VEGABAR 87 4 … 20 mA_HART - Operating Instructions User Manual

Page 67

Advertising
background image

67

11 Supplement

VEGABAR 87 • 4 … 20 mA/HART

45044-EN-131010

Version

Ambient temperature

Storage and transport temper-

ature

Version IP 68 (1 bar) with connec-

tion cable PE

-20 … +60 °C (-4 … +140 °F)

-20 … +60 °C (-4 … +140 °F)

Process conditions
Product temperature

Ʋ Suspension cable

-12 … +100 °C (+10 … +212 °F)

Ʋ Connection tube

-12 … +100 °C (+10 … +212 °F)

Mechanical stress, depending on the instrument version
Vibration resistance

4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration

with resonance)

With version with connection tube, vibration resistance 1 g. With lengths > 0.5 m (1.64 ft) the tube

must be supported in addition.
Shock resistance

100 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry

Ʋ Cable gland

M20 x 1.5

Ʋ Cable diameter (options)

5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)

Ʋ Cable entry

½ NPT

Ʋ Blind plug

M20 x 1.5; ½ NPT

Ʋ Closing cap

M20 x 1.5; ½ NPT

Plug options

Ʋ Signal circuit

Plug M12 x 1, according to ISO 4400, Harting HAN,

7/8" FF

Ʋ Indication circuit

M12 x 1 plug

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

Electromechanical data - version IP 66/IP 68 (1 bar)
Connection cable

Ʋ Configuration

four wires, one breather capillary, one suspension cable,

screen braiding, metal foil, mantle

Ʋ Wire cross-section

0.5 mm² (AWG 20)

Ʋ Wire resistance

< 0.036 Ω/m

Ʋ Tensile strength

< 1200 N (270 lbf)

Ʋ Standard length

5 m (16.4 ft)

Ʋ Max. length

180 m (590.6 ft)