4 Mounting

VEGAFLEX 81 • 4 … 20 mA/HART two-wire

41824-EN-130612

vessels, the distance to the vessel wall should be at least 500 mm

(19.7 in).
During operation, the probe must not touch any installations or the

vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to mount the

sensor in the center of the vessel, as measurement is then possible

nearly down to the lowest point of the bottom. Keep in mind that

measurement all the way down to the tip of the probe may not be pos-

sible. The exact value of the min. distance (lower dead band) is stated

in chapter "Technical data".

Fig. 7: Vessel with conical bottom

Plastic vessel/Glass vessel

The guided microwave principle requires a metal surface on the pro-

cess fitting. Therefore use in plastic vessels etc. an instrument version

with flange (from DN 50) or place a metal sheet (ø > 200 mm/8 in)

beneath the process fitting when screwing it in.
Make sure that the plate has direct contact with the process fitting.
When installing rod or cable probes in vessels without metal walls,

e.g. in plastic vessels, the measured value can be influenced by

strong electromagnetic fields (emitted interference according to

EN 61326: class A). In this case, use a probe with coaxial version.

Type of vessel