VEGA VEGAFLEX 81 4 … 20 mA_HART two-wire User Manual
Page 85
85
11 Supplement
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
41824-EN-130612
L
0
-15 mm
(-0.591
"
)
2 mm
(0.079
")
-2 mm
(-0.079
")
15 mm
(0.591
"
)
0,3 m
(11.811
")
0,15 m
(5.906
"
)
0,08 m
(3.15
")
0,05 m
(1.969
")
1
1
Fig. 51: Deviation VEGAFLEX 81 in cable version ø 2 mm (0.079 in), in oil
1 Dead band - no measurement possible in this area
L Probe length
0
-15 mm
(-0.591
"
)
2 mm
(0.079
")
-2 mm
(-0.079
")
15 mm
(0.591
"
)
0,3 m
(11.811
")
0,15 m
(5.906
"
)
0,15 m
(5.906
")
20 mm
(0.787
"
)
0,05 m
(1.969
")
1
1
L
Fig. 52: Deviation VEGAFLEX 81 in cable version ø 4 mm (0.157 in), in oil
1 Dead band - no measurement possible in this area
L Probe length
Repeatability
≤ ±1 mm
Variables influencing measurement accuracy
Specifications for the digital measured value
Temperature drift - Digital output
±3 mm/10 K relating to the max. measuring range or
max. 10 mm (0.394 in)