VEGA VEGAFLEX 81 4 … 20 mA_HART two-wire User Manual

Page 85

Advertising
background image

85

11 Supplement

VEGAFLEX 81 • 4 … 20 mA/HART two-wire

41824-EN-130612

L

0

-15 mm

(-0.591

"

)

2 mm

(0.079

")

-2 mm

(-0.079

")

15 mm

(0.591

"

)

0,3 m

(11.811

")

0,15 m

(5.906

"

)

0,08 m

(3.15

")

0,05 m

(1.969

")

1

1

Fig. 51: Deviation VEGAFLEX 81 in cable version ø 2 mm (0.079 in), in oil
1 Dead band - no measurement possible in this area

L Probe length

0

-15 mm

(-0.591

"

)

2 mm

(0.079

")

-2 mm

(-0.079

")

15 mm

(0.591

"

)

0,3 m

(11.811

")

0,15 m

(5.906

"

)

0,15 m

(5.906

")

20 mm

(0.787

"

)

0,05 m

(1.969

")

1

1

L

Fig. 52: Deviation VEGAFLEX 81 in cable version ø 4 mm (0.157 in), in oil
1 Dead band - no measurement possible in this area

L Probe length

Repeatability

≤ ±1 mm

Variables influencing measurement accuracy
Specifications for the digital measured value
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm (0.394 in)