Code
Text mes-
sage

Cause

Rectification

F

125

Unpermissible
electronics
temperature

l Temperature of the elect-

ronics in the non-specified
section

l Check ambient temperatu-

re

l Isolate electronics
l Use instrument with higher

temperature range

F

260

Error in the
calibration

l Error in the calibration car-

ried out in the factory

l Error in the EEPROM

l Exchange of the electro-

nics

l Send instrument for repair

F

261

Error in the
configuration

l Error during setup
l False signal suppression

faulty

l Error when carrying out a

reset

l Repeat setup
l Repeat reset

F

264

Installation/S-
etup error

l Adjustment not within the

vessel height/measuring
range

l Max. measuring range of

the instrument not suffici-
ent

l Check or correct installa-

tion and/or parameter ad-
justment

l Use an instrument with

bigger measuring range

F

265

Measurement
function dis-
turbed

l Sensor does no longer

carry out a measurement

l Operating voltage too low

l Check operating voltage
l Carry out a reset
l Separate operating voltage

briefly

The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.

Code
Text mes-
sage

Cause

Rectification

C

700

Simulation

l Simulation active

l Finish simulation
l Wait for the automatic end

after 60 mins.

The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.

Function check

Out of specification