VEGA VEGABAR 82 HART and accumulator pack - Operating Instructions User Manual

Page 37

Advertising
background image

37

6 Set up with the display and adjustment module

VEGABAR 82 • HART and accumulator pack

45029-EN-131107

Reset - Diagnosis

Menu item

Parameter

Default value

Sensor status

-

Peak value

Pressure

Actual measured value

Temperature

Actual temperature values from meas-

uring cell, electronics

Simulation

Process pressure

Reset - Additional settings

Menu item

Parameter

Default value

PIN

0000

Date/Time

Actual date/Actual time

Copy in-

strument

settings
Special pa-

rameters

No reset

Scaling

Scaling size

Volume in l

Scaling format

0 % corresponds to 0 l
100 % corresponds to 0 l

Current out-

put

Current output - Size

Lin. percent - Level

Current output - adjustment

0 … 100 % correspond to

4 … 20 mA

HART mode

Address 0

The instrument settings are copied with this function. The following

functions are available:

Read from sensor: Read data from sensor and store into the

display and adjustment module

Write into sensor: Store data from the display and adjustment

module back to the sensor

The following data or settings for adjustment of the display and ad-

justment module are saved:

All data of the menu "Setup" and "Display"

In the menu "Additional adjustments" the items "Reset, Date/Time"

The user-programmable linearization curve

The copied data are permanently saved in an EEPROM memory in

the display and adjustment module and remain there even in case of

power failure. From there, they can be written into one or more sen-

sors or kept as backup for a possible electronics exchange.

Additional adjustments -

Copy instrument settings

Advertising