VEGA VEGABAR 82 HART and accumulator pack - Operating Instructions User Manual

Page 61

Advertising
background image

61

11 Supplement

VEGABAR 82 • HART and accumulator pack

45029-EN-131107

Integrated clock
Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Rate deviation max.

10.5 min/year

Measurement electronics temerature
Resolution

1 °C (1.8 °F)

Accuracy

±1 °C (1.8 °F)

External battery charger
Mains voltage

100 … 240 V AC

Output voltage

24 V DC

Output current (shortcircuit proof) max. 500 mA
Load current limitation

70 mA

DC plug (inside plus, outside minus)

2.1 mm

Integrated accumulator
Type

Lithium ions

Voltage

14.8 V

Accumulator capacity

4.7 Wh

Charging period from 0 % to 100 %

approx.

4 h

Operating time after 10 min. charging of

0 %

> 3 h

Operating time in mode 4 (sensor per-

manently on) with completely charged

accumulator

> 60 h

Temperature range

Ʋ Charge accumulator

0 … +45° C (+32 … +167 °F)

Ʋ Accumulator operation

-20 … +60° C (-4 … +140 °F)

Temperature derating accumulator capacity

Ʋ +25° C (+77 °F)

100 %

Ʋ -10° C (+14 °F)

50 %

Potential connections in the instrument
Electronics

Not non-floating

Ground terminal

Galvanically connected with the process fitting

Electrical protective measures

Housing material

Version

IP-protection class

NEMA protection

Plastic

Single chamber

IP 66/IP 67

NEMA 4X/6P

Double chamber

IP 66/IP 67

NEMA 4X

Advertising