VEGA VEGAFLEX 86 4 … 20 mA_HART - four-wire User Manual

Page 56

Advertising
background image

56

7 Setup with PACTware

VEGAFLEX 86 • 4 … 20 mA/HART - four-wire

41845-EN-130314

Adjustment for the level and interface measurement

If you have selected interface measurement in the previous menu,

then you can enter the values for the min. and max. adjustment of

level and interface or accept the values of the level measurement. The

entered value refers to the distance from the sealing surface of the

process fitting (sensor reference plane) to the total level or interface.

Linearization is required if the measured value should be outputted

in proportion to the volume and not the level. The linearization acts

identically on the level and the interface measurement. You can find

further linearization types in the extended adjustment.
If you have a non-linear vessel, you can select here the respective

linearization curve.

Linear

Spherical tank

Horizontal cylindrical tank

You must enter the following vessel dimensions with non-linear condi-

tions:

Height of the socket h

Vessel height D

In this window you can adjust the output signal. When the function is

inactive, you can change the settings via the "Extended adjustment".

These settings allow you to optimize the sensor. With them you can

compare a sounded distance with the indicated value and correct it, if

necessary.
Probe immersed in the liquid (covered)

Select whether or not the probe is immersed in the medium.
Measured distance to the medium

If the probe is immersed in the medium, you can enter here the meas-

ured distance to the medium.
Displayed distance correct?

Is the displayed distance value correct? If you have the possibility, you

can enter here the sounded distance to the medium.
False signal suppression

With this function you can carry out the automatic false signal sup-

pression. We recommend carrying out false signal suppression in any

case.

When the setup of the instrument is finished, additional settings can

be made. These are various backups and the locking of the instru-

ment against unauthorised or inadvertent adjustment.
Prepare a backup file of the instrument parameter adjustment?

For backup purposes, the current parameter adjustment of the instru-

ment is stored in a file. You can use this file later on to restore the

Step 4 of 7

Linearization

Step 5 of 7

Setting - Output signal

Step 6 of 7

Sensor optimization

Step 7 of 7

Additional settings

Advertising