6 saving the parameter adjustment data – VEGA VEGAFLEX 81 Modbus and Levelmaster protocol User Manual

Page 47

Advertising
background image

47

6 Set up with the display and adjustment module

VEGAFLEX 81 • Modbus and Levelmaster protocol

41828-EN-130314

The default setting is "Analogue current output" and the address 00.

In this menu, you read out the instrument name and the instrument

serial number:

In this menu item, the hardware and software version of the sensor is

displayed.

In this menu item, the date of factory calibration of the sensor as well

as the date of the last change of sensor parameters are displayed via

the display and adjustment module or via the PC.

In this menu item, the features of the sensor such as approval, pro-

cess fitting, seal, measuring range, electronics, housing and others

are displayed.

6.6 Saving the parameter adjustment data

We recommended noting the adjusted data, e.g. in this operating

instructions manual, and archiving them afterwards. They are thus

available for multiple use or service purposes.
If the instrument is equipped with a display and adjustment module,

the data in the sensor can be saved in the display and adjustment

module. The procedure is described in the operating instructions

manual "Display and adjustment module" in the menu item "Copy

sensor data". The data remain there permanently even if the sensor

power supply fails.
The following data or settings for adjustment of the display and ad-

justment module are saved:

Info - Instrument name

Info - Instrument version

Info - Factory calibration

date

Info - Sensor character-

istics

Advertising