8 Diagnosis, asset management and service

VEGABAR 86 • Modbus and Levelmaster protocol

46296-EN-131107

8.6 Exchange process assembly with version

IP 68 (25 bar)

With version IP 68 (25 bar), the user can exchange the process as-

sembly on site. Connection cable and external housing can be kept.
Required tools:

Hexagon socket wrench, size 2

Caution:

The exchange may only be carried out in the complete absence of

line voltage.
In Ex applications, only a replacement part with appropriate Ex ap-

proval may be used.
Proceed as follows when carrying out the exchange:
1. Losen the fixing screw with the hexagon socket wrench
2. Carefully detach the cable assembly from the process assembly

1

2

5

3

4

Fig. 26: VEGABAR 86 in IP 68 version, 25 bar and lateral cable outlet, external

housing
1 Process assembly

2 Plug connector

3 Cable assembly

4 Connection cable

5 External housing

3. Loosen the plug connector
4. Mount the new process assembly on the measuring point
5. Plug the connector back in