VEGA VEGABAR 86 Modbus and Levelmaster protocol - Operating Instructions User Manual

Page 57

Advertising
background image

57

10 Supplement

VEGABAR 86 • Modbus and Levelmaster protocol

46296-EN-131107

Deviation (according to IEC 60770)
Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to the

analogue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal

measuring range/set span.
The specified values correspond to the value F

Kl

in chapter "Calculation of the total deviation".

Accuracy class

Non-linearity, hysteresis and re-

peatability with TD 1 : 1 up to 5 : 1

Non-linearity, hysteresis and re-

peatability with 5 : 1

0.1 %

< 0.1 %

< 0.02 % x TD

Influence of the product or ambient temperature
Thermal change zero signal and output span
Turn down (TD) is the relation nominal measuring range/adjusted span.
Ceramic measuring cell - Standard

Medium or ambient temperature T
-40 … 0 °C (-40 … +32 °F)

< {0.15 % + 0.015 %/K x (-T)} x {0.5+0.5 x TD}

0 … 40 °C (+32 … +104 °F)

< {0.0075 %/K x (T - 20 °C)} x {0.5+0.5 x TD}

40 … 100 °C (+104 … +212 °F)

< {0.15 % } x {0.5+0.5 x TD}

100 … 120 °C (+212 … +248 °F)

< {0.15 % + 0.005 %/K x (T - 100 °C)} x {0.5+0.5 x TD}

Ceramic measuring cell - climate compensated
Compensated temperature range

0 … +100 °C (+32 … +212 °F)

Non-compensated temperature range

-40 … 0 °C (-40 … +32 °F), -100 … 130 °C

(212 … +276 °F)

Nominal measuring range in bar/

kPa

Nominal measuring

range in psig

In the compensated

temperature range

In the non-compen-

sated temperature

range

0 … 10 bar/0 … 1000 kPa

0 … 150 psig

< {0.05 % + 0.1 x

(0.5+0.5 x TD)}

< {0.1 % + 0.15 x

(0.5+0.5 x TD)}

0 … 25 bar/0 … 2500 kPa

0 … 350 psig

0 … 1 bar/0 … 100 kPa

0 … 15 psig

< {0.1 % + 0.2 x

(0.5+0.5 x TD)}

< {0.1 % + 0.3 x

(0.5+0.5 x TD)}

0 … 2.5 bar/0 … 250 kPa

0 … 35 psig

0 … 0.4 bar/0 … 40 kPa

0 … 6 psig

< {0.15 % + 0.4 x

(0.5+0.5 x TD)}

< {0.15 % + 0.6 x

(0.5+0.5 x TD)}

>>>Long-term stability (according to DIN 16086 and IEC 60770-1)
Specifications refer to the set span. Turn down (TD) is the ratio: nominal measuring range/set span.
Long-term drift of the zero signal

Time period

All measuring ranges

Measuring range

0 … +0.025 bar/0 … +2.5 kPa

One year

< 0.05 % x TD

< 0.1 % x TD

Five years

< 0.1 % x TD

< 0.2 % x TD

Ten years

< 0.2 % x TD

< 0.4 % x TD

Advertising