VEGA VEGAPULS 61 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA__HART four-wire User Manual

Page 33

Advertising
background image

B

asic adjustment

D

isplay

D

iagnostics

S

ervice

I

nfo

2

S

elect the menu item "Basic adjustment" with [->] and confirm with

[OK]

. N

ow the menu item "Min. adjustment" is displayed.

M

in. adjustment

0

.00 %

=

5

.000 m(d)

4.000

m(d)

3

P

repare the % value for editing with [OK] and set the cursor to the

requested position with [->]. Set the requested percentage value
with [+] and save with [OK]. The cursor jumps now to the distance
value.

4

E

nter the suitable distance value in m for the empty vessel (e.g.

distance from the sensor to the vessel bottom) corresponding to
the percentage value.

5

S

ave the settings with [OK] and move to "Max. adjustment" with [-

>

]

.

P

roceed as follows:

M

ax. adjustment

100

.00 %

=

1

.000 m(d)

2.000

m(d)

1

P

repare the % value for editing with [OK] and set the cursor to the

requested position with [->]. Set the requested percentage value
with [+] and save with [OK]. The cursor jumps now to the distance
value.

2

E

nter the appropriate distance value in m (corresponding to the

percentage value) for the full vessel. Keep in mind that the max.
level must lie below the dead band.

3

S

ave the settings with [OK] and move to "Medium selection" with

[->]

.

E

ach product has different reflective properties. In addition, there are

various interfering factors which have to be taken into account:
agitated product surfaces and foam generation (with liquids); dust
generation, material cones and echoes from the vessel wall (with
solids). To adapt the sensor to these different conditions, you should

rst select "Liquid" or "Solid".

Carrying out max. ad-
justment

Medium selection

VEGAPULS

61 • 4 … 20

mA/HART four-wire

33

6 S

et up with the indicating and adjustment module PLICSCOM

28439

-EN