Prokit's Industries MT-1620 User Manual

Page 28

Advertising
background image

4.9 DC VOLTAGE MEASURING

WARNING

You can’t input the voltage which more than

600V DC, it’s possible to show higher voltage,

but it’s may destroy the inner circuit.

Pay attention not to get an electric shock

when measuring high voltage.

4.9.1 Connect the black test lead to the COM jack

and the red test lead to the V jack.

4.9.2 Set the transform switch at the V range

position.

4.9.3 Put down the "FUNC." to enter the DC

measurement. Auto range or manual range
can be transformed by putting the “RANGE”.

4.9.4 Connect test leads across the source or load

under measurement.

4.9.5 You can get reading from LCD. The polarity

of the red lead connection will be indicated
along with the voltage value.

NOTE:

 When only the figure ‘OL’ is displayed, it

indicates overrange situation and the higher
range has to be selected.

- 27 -