Figure 29: Shrinkage stress of various composites Source: Gc corporation

the volumetric shrinkage was measured

following iSO Draft Date: 2007-07-10

(Dentistry - Polymerization shrinkage of

filling materials) specifications.

this study demonstrated that
the shrinkage stress generated
by G-ænial is amongst the
lowest of the tested
composites.

6.3 three – body wear resistance

Wear is the loss of material resulting from removal of the material through the contact of two or
more materials. the three-body wear test simulates wear in the oral cavity using a slurry of Pmma
and glycerol as the intermediate abrasive agent and an acrylic plate as the opposing material.

Figure 30: three-Body wear resistance test set-up

1mm

2mm

2mm

2mm

7mm

Sample holder

Composite slurry

Acrylic plate

Shrinkage Stress (n)