G-ænial Anterior & Posterior Technical Manual

table 7: G-ænial anterior: irradiation time and effective depth of cure

irradiation time

Plasma arc (2000 mW/cm²)

Gc G-light (1200 mW/cm²)

halogen / leD (700 mW/cm²)

Shade

3 sec.
10 sec.
20 sec.

6 sec.
20 sec.
40 sec.

te, ie, Je, Se, cVe

3.0 mm

3.5 mm

a1, a2, B1, B2, XBW, BW, ae

2.5 mm

3.0 mm

a3, B3

2.0 mm

3.0 mm

a3.5, a4, c3, aO2, aO3, aO4, cV, cVD

1.5 mm

2.5 mm

table 8: G-ænial Posterior: irradiation time and effective depth of cure

irradiation time

Plasma arc (2000 mW/cm²)

Gc G-light (1200 mW/cm²)

halogen / leD (700 mW/cm²)

Shade

3 sec.
10 sec.
20 sec.

6 sec.
20 sec.
40 sec.

P-a1, P-a2, P-Je, P-ie

2.5 mm

3.0 mm

P-a3, P-a3.5

2.0 mm

3.0 mm

note :
1. material should be placed and light cured in layers. For maximum layer thickness, please consult

the tables.

2. lower light intensity may cause insufficient curing or discoloration of the material.

7. Finishing and Polishing

Finish and polish using diamond burs, polishing points and discs. to obtain a high gloss, polishing
pastes can be used.

STORAGE

Store in a cool and dark place (4 – 25°c / 39.2 – 77.0°F) away from high temperatures or direct
sunlight. (Shelf life: 3 years from date of manufacture)

CAUTION

1. in case of contact with oral tissue or skin, remove immediately with cotton or a sponge soaked

in alcohol. Flush with water.

2. in case of contact with eyes, flush immediately with water and seek medical attention.
3. take care to avoid ingestion of the material.
4. Wear plastic or rubber gloves during operation to avoid direct contact with air inhibited resin

layers in order to prevent possible sensitivity.

5. For infection control reasons, unitips are for single use only.
6. Wear protective eye glasses during light curing.
7. When polishing the polymerised material, use a dust collector and wear a dust mask to avoid

inhalation of cutting dust.

8. Do not mix with other similar products.
9. avoid getting material on clothing.
10. in case of contact with unintended areas of tooth or prosthetic appliances, remove with

instrument, sponge or cotton pellet before light curing.

11. Do not use G-ænial in combination with eugenol containing materials as eugenol may hinder

G-ænial from setting.

12. all shades except for the outside special shade te are radiopaque.