Analytical Industries AII-4001 Helium Oxygen Trimix Analyzer User Manual

Page 2

Advertising
background image

1

Table of Contents





















1 Introduction

Congratulations you have purchased the most advanced portable gas analyzer
available. The AII-4001 provides unmatched sophisticated performance in a
simple to use package, and, provides users with significant advantages:

Precisely analyze dive gas mixes in any environment without charts. Proprietary
algorithms* and sensors provide extremely accurate one touch calibration.
Note: Failure to account for environmental factors can produce up to 6.7%
error.

Oxygen is compensated for temperature, barometric pressure and RH.
Helium is zeroed when the oxygen is calibrated with air or 100% oxygen.

Algorithms* compensate the Helium value for differences in the thermal con-
ductivity of Oxygen and Nitrogen enabling the AII-4001 to precisely measure
any combination of Helium, Oxygen and Nitrogen.

The large backlit LCD is easily read even in low light condition and provides

additional information to the user:

Temperature, barometric pressure and RH values of the sample gas.

For added safety, the Maximum Operating Depth (MOD) PO2 @ 1.4 ATA of
the gas mix being sampled

Imperial or Metric units at a touch

Rechargeable battery powers the AII-4001 for 16 hours of continuous use.

9 VDC power adapter 110/220 VAC is included, charges battery in 2 hours
The adapter can safely remain connected for bench top operation

Optional 9 VDC cigarette lighter adapter is available

Automatically conserves and protects battery life.

The backlight turns off after 30 seconds and

The AII-4001 completely powers itself off after 15 minutes of non-use
The automatic turn-off features are disabled when externally powered

The analyzer turns off 2-3 hours after the LOW BATTERY warning is displayed
to prevent complete discharge and permanent damage to the battery

The AII-4001 and accessories are packaged in a rugged waterproof plastic enclosure de-

signed specifically for hostile diving conditions. The unmistakable bright red color stands

out in the shop, on the boat or in the water.

* Patent pending


1.1 Contents
Open the red case containing the AII-4001 and confirm you have received the

standard accessories and any optional equipment ordered:

1) AII-4001 Helium Oxygen Trimix Analyzer

2) Flow Manifold (A-3582)

3) Adapter, Dome to 1/8” Tube (A-3609)

4) Tubing, 1/8” x 3 ft. (TUBE-1018)

5) Adapter, 9 VDC 110/220 VAC (PWRS-1003-1)

6) Owner’s Manual







2

1 Introduction

1

2 Quality

Control

Certification 3

3

General Care & Safety

5

4 Specifications

6

5 Operation

7

5.1 Controls

7

5.2 Power

8

5.3 O

2

Calibration & He Zero

9

5.4 Analysis

10

6 Maintenance

11

7

Replacement Parts & Accessories

12

8 Troubleshooting

13

9 Warranty

14

10

MSDS & Disposal

15

Advertising