8 Troubleshooting

9 Warranty


Coverage

Under normal operating conditions, the analyzer and sensors are warranted to
be free of defects in materials and workmanship for the period specified in the
current published specifications. To make a warranty claim, you must return
the item properly packaged and postage prepaid to:

Analytical Industries Inc.
2855 Metropolitan Place
Pomona, Ca 91767 USA


Analytical Industries in their sole discretion shall determine the nature of the
defect. If the item is determined to be eligible for warranty we will repair it or,
at our option, replace it at no charge to you. If we choose to repair your item,
we may use new or reconditioned replacement parts. If we choose to replace
your item, we may replace it with new or reconditioned components of the
same or upgraded design. This is the only warranty we will give and it sets
forth all our responsibilities, there are no other express or implied warranties.


The warranty period begins with the date of shipment from Analytical Indus-
tries and is limited to the first customer who submits a claim for a given serial
number which must be in place and readable to be eligible for warranty. Under
no circumstances will the warranty extend to more than one customer or be-
yond the warranty period.

Exclusions
This warranty does not cover normal wear and tear; corrosion; damage while
in transit; damage resulting from misuse or abuse; lack of proper maintenance;
unauthorized repair or modification of the analyzer; fire; flood; explosion or
other failure to follow the Owner’s Manual.

Limitations
Analytical Industries shall not liable for losses or damages of any kind; loss of
use of the analyzer; incidental or consequential losses or damages; damages

resulting from alterations, misuse, abuse, lack of proper maintenance; unau-
thorized repair or modification of the analyzer.

Service
Telephone 909-392-6900, fax 909-392-3665 or e-mail [email protected] be-
tween 8:00am and 5:00pm PST Monday thru Thursday or before 12:00pm on
Friday. Trained technicians will assist you in diagnosing the problem and deter-
mining the appropriate course of action.

14

Symptom

Possible Cause

Recommendation

Display blank

15 min auto turn off feature

Battery protection auto turn off

Bad connection

Other component failure

Press red POWER key

Charge analyzer for 2 hrs

Check battery connection

Return to factory

Display dark

Auto turn off feature

Exposure to high temperature

Press yellow BACK LIGHT key

Move the unit to cooler place

Display segment missing LCD component failure

Return to factory

Display reflects empty

battery symbol

LOW BATTERY warning

Charge battery

Readings unstable

Bad electrical connection

Ignore LOW BATTERY warning

Backpressure on sensor

RF/Electro-mag interference

O

2

sensor nearly expired

He sensor faulty

Check connections

Charge or replace battery

Check pressure, flow and vent

(remove any restriction)

Move unit away from source

Replace O

2

sensor

Check with known He % mix

Replace He sensor

Reading drifts

Leak in gas path connections

Liquid covering sensing areas

Ignore LOW BATTERY warning

Temperature change >10⁰C

O

2

sensor nearly expired

He sensor faulty

Check connections

Remove liquid

Replace O

2

sensor

Charge or replace battery

Stabilize 30 min at new temp

Replace O

2

sensor

Check with known He % mix

Replace He sensor

Reading zero O

2

Bad

electrical

connection

Liquid covering sensing area

No oxygen present

Shock from being dropped

O

2

sensor expired

Check O

2

sensor connection

Remove liquid

Expose to air

Replace PCB or O

2

sensor

Replace O

2

sensor

Reading +/- He in air

He zero required

He sensor fault

Perform CAL O

2

ZERO He

Replace He sensor

O

2

Calibration failed

No oxygen present

O

2

sensor nearly expired

Shock from being dropped

Expose to air, repeat

Replace O

2

sensor

Replace PCB, sensors

He Zero failed

He sensor disconnected

He sensor faulty

Check He sensor connection

Check with known He % mix