10 msds - material safety data sheet, Disposal – Analytical Industries AII-4001 Helium Oxygen Trimix Analyzer User Manual

Page 9

Advertising
background image

15

10 MSDS - Material Safety Data Sheet

Disposal

O2 sensor and battery should be disposed of in accordance with local regula-
tions for batteries.



WEEE regulations prohibit electronic products including the He-

lium and environmental sensors from being placed in household
trash bins.

Electronic products including the Helium and environmental sen-
sors should be disposed of in accordance with local regulations.

16

Product name Electrochemical Galvanic Fuel Cell Oxygen Sensor

Exposure

Sealed device with protective coverings, normally no hazard

Ingredients

Carcinogens - none; Potassium Hydroxide (KOH), Lead (Pb)

Properties

Completely soluble in H

2

O; evaporation similar to H

2

O

Flash Points

Not applicable, non-flammable

Reactivity Stable;

avoid mixing strong acids, emits fumes when heated

Health Hazard KOH entry via ingestion - harmful or fatal if swallowed; eye -

corrosive, possible loss of vision; and, skin contact - corrosive,
possible chemical burn. Liquid inhalation is unlikely.
Lead - known to cause birth defects, but contact unlikely

Symptoms

Eye contact - burning sensation; skin contact - slick feeling

Protection

Ventilation - none; eye - safety glasses; hands - gloves

Precautions

Do not remove Teflon and PCB coverings; do not probe with
sharp objects; avoid contact with eyes, skin and clothing.

Action KOH
Leak

Use rubber gloves, safety glasses and H

2

O and flush all sur-

faces repeatedly with liberal amounts of H

2

O

Advantages of the most advanced portable
analyzer available . . .

Precisely analyze dive gas mixes in any environment without charts.

Proprietary algorithms* and sensors provide extremely accurate one touch calibration.

Note: Failure to account for environmental factors can produce up to 6.7% error.

Oxygen value is compensated for temperature, barometric pressure and RH.

Helium value is zeroed when the oxygen is calibrated with air or 100% oxygen.

Algorithms* compensate the Helium value for differences in the thermal conductivity of

Oxygen and Nitrogen enabling the AII-4001 to precisely measure any combination of

Helium, Oxygen and Nitrogen.

The large backlit LCD is easily read even in low light condition and provides

additional information to the user:

Temperature, barometric pressure and RH values of the sample gas.
For added safety, the Maximum Operating Depth (MOD) PO2 @ 1.4 ATA of

the gas mix being sampled

Imperial or Metric units at a touch

Rechargeable battery powers the AII-4001 for 16 hours of continuous use.

9 VDC power adapter 110/220 VAC is included, charges battery in 2 hours

The adapter can safely remain connected for bench top operation

Optional 9 VDC cigarette lighter adapter is available

Automatically conserves and protects battery life.

The backlight turns off after 30 seconds and
The AII-4001 completely powers itself off after 15 minutes of non-use

The automatic turn-off features are disabled when externally powered

The analyzer turns off 2-3 hours after the LOW BATTERY warning is displayed
to prevent complete discharge and permanent damage to the battery

The AII-4001 and accessories are packaged in a rugged waterproof

plastic enclosure designed specifically for hostile diving conditions.

The unmistakable bright red color stands out in the shop, on the

boat or in the water.

* Patent pending