2 quality control certification – Analytical Industries AII-4001 Helium Oxygen Trimix Analyzer User Manual

Page 3

Advertising
background image

2 Quality Control Certification

3

4

Date: ________ Customer _____________________________ Order _____________

Model:

AII-4001 Helium Oxygen Analyzer

S/N _____________

Sensors:

AII-11-75D Oxygen Sensor

S/N _____________

AII-41-100 Helium Sensor

S/N _____________

AII-E3 Environmental Sensor Module

S/N _____________

Configuration: A-1185 PCB Assembly, Main Software Ver ………... _____________

Accessories:

A-3582 Flow Manifold ……………………………………………………….. ______

A-3609 Adapter, Dome to 1/8” Tube …...…………………………….. ______

TUBE-1018 Tubing 1/8” x 3 ft …………………………………………….. ______

PWRS-1003-1 Charger 9VDC 110/220 VAC Adapter ……………... ______

Instruction Manual ………………………………………………………………. ______

Testing:

Mechanical: all screws, sensors and connectors secured …………. ______

Labels: 3 labels applied to case …………………………………………….. ______

LCD display 3-1/2 digits ……………………………………………………….. ______

Backlight: on for 3 seconds after power-up ……………….…………...

_____

on for 15 seconds after key press ………………………….. ______

Battery icon: “LOW BATTERY” displays in reverse background

in lower right hand corner when the battery icon

displays < 30% charge ………………………..…………. ______

displays solid black when fully charged ….……….. ______

Charging battery: “CHARGING” displays in lower right hand

corner when the charge plug is connected

to the analyzer and the adapter plugged into
a 110/220 VAC outlet ……………………...……..…

______

Test (cont’d)

Temperature: within 2⁰F of EXTECH RH300 Digital Psychrometer
at ambient conditions ……………………………………….. ______

Relative Humidity: within 3% of EXTECH RH300 Digital

Psychrometer at ambient conditions ……….….. ______

Barometric Pressure: within 1 KPa of Accu Cal Plus Digital

Precision Test Gauge ……………………………... ______

Calibration O2 Sensor: +10-30% FS in ambient air ………………..… _______

>98% in 100% oxygen ……………………….. ______

Zero He Sensor: +1% FS in 100% oxygen ………………………………. ______

>98% in 100% helium span gas …..……………….. ______

O2 Sensor: within 0.5% of ambient air calibration value …………... ______

Overall Appearance: clean, no physical defects ………..………………. ______

Part Number & Description _Qty_

Options:

PWRS-1019 Adapter, 9VDC Cigarette Lighter …………………………… ______

A-3673 Adapter, BC with Flow Resistor to 1/8” Tube ………...……. ______

A-3677 Adapter, DIN to 1/8” Tube ………………………………………... ______

A-3678 Adapter, A-Yoke to 1/8” Tube ……………………………………. ______

Notes:

Delivery: