2 measuring the ph value, Measuring the ph value – Xylem handylab pH 11 User Manual

Page 17

Advertising
background image

handylab pH 11

Operation

17

ba75366e04

01/2011

4.2.2

Measuring the pH value

Setting the

resolution

The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the <M> key
while pressing the <RUN/ENTER> key.

AutoRead

(drift control)

The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.

For identical measurement conditions, the following criteria apply:

1

Perform the preparatory activities according to section 4.2.1.

2

Immerse the pH electrode in the test sample.

3

Press the <M> key until pH appears on the status display. The
pH value appears on the display.

mV/pH

mol/L

8

Auto Store

TempError

RCL

LoBat

AutoCal DIN

CalError

ARng AR

AutoCal TEC ConCal

Time

Baud

Day.Month No.

Ident

Year

1

U

%

°F

S

mg/L

% ppm

ISE

8

4

2

2

99

6

°C

TP

pH

Reproducibility

Response time

Better than 0.02

> 30 seconds

1

Call up the pH measuring mode with <M> and <

> <> .

2

Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).

3

Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.

4

If necessary, start the next AutoRead measurement with
<RUN/ENTER>.

5

To terminate the AutoRead function: Press the <AR> key.